The shield for the Single-Tube Station on the DxH instrument may interfere with sample aspiration in the open vial position.
Source facts
Action
The firm notified its consignees of the recall on 11/03/2017 by letter. The letter stated the issue and provided the following instructions:
"To avoid a biohazardous spill, do not use the right (green) position of the Single-Tube Station for sample aspiration until the shield is removed.
You may remove the shield following the instructions in Attachment 1 or contact your local Beckman Coulter Representative for further assistance."
Initiated
2017-11-03
Root cause
Device Design
Terminated
2018-06-19
Distribution
CA, AR, HI, CA, OH, WA, MD, Canada, Japan
Recall number
Z-0238-2018
Recall status
Terminated
CDRH record ID
159891
FDA record created
2017-12-13
Recall event number
78611