Z-0238-2022: Philips Azurion systems with software releases 2.1(L1) and 2.1(L2)

Recall record Z-0238-2022 for Philips Healthcare. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0238-2022
Company in FDA record
Philips Healthcare
FDA record creation date
Nov 24, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

3D-RA is a reconstruction software product that can be used with the Philips Azurin system. This product extends the functionality of compatible X-ray equipment with the capability of 3D X-ray imaging. Before starting a 3D-RA scan, the detector of the Philips Azurion system must be positioned in portrait or landscape orientation with an angle of 0 or +/-90 degrees relative to the 3D scan direction. Philips has discovered that Azurion releases 2.1(L1) and 2.1(L2) allow creation of a 3D-RA scan with a deviating detector orientation of up to 3 degrees from the exact portrait or landscape orientation (i.e., at 0 or +/-90 degrees), relative to the 3D scan direction. 3D-RA scans with a deviating detector orientation of ore than 1.0 degree from the exact portrait or landscape orientation (i.e., at 0 or +/-90 degrees) cannot be reconstructed with the 3D-RA reconstruction software.

Source facts

Action
Philips will inform customers of the issue via an Electronic Product Radiation Correction notification. The letter identified the affected product, problem and actions to be taken. Philips Field Service Engineers will schedule an appointment with customers to install the software update. For questions contact your Philips representative at 1-800-722-9377.
Initiated
2021-09-24
Root cause
Radiation Control for Health and Safety Act
Distribution
US Nationwide distribution
Recall number
Z-0238-2022
Recall status
Open, Classified
CDRH record ID
190291
FDA record created
2021-11-24
Recall event number
89020

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗