Open original FDA record ↗Recall record
- FDA identifier
- Z-0239-2019
- Company in FDA record
- Philips Medical Systems Gmbh, DMC
- FDA record creation date
- Oct 24, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Stitching run on SkyPlate detector interrupted intermittently; Stitching with SkyPlate aborts after first image; System software CD contains wrong baseline
Source facts
- Action
- Philips issued letter dated 9/10/18 to advise users of the problem. Philips will install a system software version upgrade in affected systems, which will eliminate this issue. A Philips Service Engineer will contact account when the Field Action Kit is available to be implemented. Should you need to communicate with Philips with regard to this program, please reference Field Change Order 71200177.
- Initiated
- 2018-09-19
- Root cause
- Software design
- Terminated
- 2023-08-10
- Distribution
- Worldwide distribution - US Nationwide and countries of Canada Argentina Australia Austria Bangladesh Belgium Bolivia Brazil Chile China Colombia Czech Republic Denmark Egypt Finland France French Guiana Germany Guadeloupe Hong Kong Hungary India Indonesia Iran Israel Italy Japan Jordan Korea, Republic of Kuwait Kyrgyzstan Latvia Luxembourg Maldives Martinique Mexico Netherlands New Zealand Norway Panama Peru Poland Portugal Qatar Romania Russian Federation Saudi Arabia Singapore Slovenia South Africa Spain Sweden Switzerland Taiwan Thailand Turkey United Arab Emirates United Kingdom
- Recall number
- Z-0239-2019
- Recall status
- Terminated
- CDRH record ID
- 168182
- FDA record created
- 2018-10-24
- Recall event number
- 81310