Open original FDA record ↗Recall record
- FDA identifier
- Z-0240-2008
- Company in FDA record
- Hill-Rom, Inc.
- FDA classification date
- Jul 24, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
False Latch/Failure to Latch-The side rail may not latch, or the latch may not fully engage, resulting in the siderail being able to be lowered without pulling on the release handle.
Source facts
- Action
- The firm notified consignees via an Urgent Device Correction letter dated 11/12/07 informing them of the problem and informed them they would be receiving upgrade kits which they could have their qualified personnel install themselves following video instructions also provided, or they could contact Hill-Rom for assistance.
- Initiated
- 2006-11-12
- Root cause
- Device Design
- Terminated
- 2010-10-22
- Distribution
- Worldwide, including USA, Argentina, Australia, Brazil, Canada, Dominican Republic, Emirate of Sharjah, England, Greece, Indonesia, Japan, Jordan, Kuwait, Lebanon, Malaysia, Mexico, Oman, Pakistan, Philippines, Qatar, Saudi Arabia, Selangor, Singapore, South Korea, Spain, Taiwan, Thailand, Turkey, United Arab Emirates and Venezuela.
- Recall number
- Z-0240-2008
- Recall status
- Terminated
- CDRH record ID
- 53251
- FDA classified
- 2007-07-24
- Recall event number
- 38256