The Gamma Nail may have been manufactured with an incorrect lag screw hole angulation.
Catalog number 3225-0360S, lot code K949396 was manufactured with a lag screw hole angulation of 120 degrees instead of 125 degrees. Catalog number 3220-0320S, lot code K949408 was manufactured with a lag screw hole angulation of 125 degrees instead of 120 degrees.
Source facts
Action
Notification letters and Product Accountability forms were sent via FedEx on 11/22/2004 with return receipt.
Initiated
2004-11-22
Root cause
Other
Terminated
2006-10-03
Distribution
The product was sent to Stryker Orthopaedics branches and agents throughout the US. There was one hospital that was a direct account.
Recall number
Z-0241-06
Recall status
Terminated
CDRH record ID
38869
FDA classified
2005-12-01
Recall event number
31985
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.