Z-0241-2026: Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system

Recall record Z-0241-2026 for PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0241-2026
Company in FDA record
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
FDA classification date
Oct 23, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Over time some systems may experience loss of imaging functionality and/or poor image quality due to potential corrosion of the Image Intensifier Television (IITV) control board.

Source facts

Action
On September 29, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: 1. Circulate this notice to all users and follow the instructions in the letter. 2. Affected systems may continue to be used in accordance with their intended use and Instructions for Use (IFU). 3. Ensure the letter is in a place likely to be seen/viewed. 4. In case the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction letter to that organization and inform Philips about this transfer through your local Philips representative. 5. If you experience the issue described in this letter, please report it to your local Philips representative. Actions by Philips: Starting in October 2025, Philips will replace the 11W control board of the affected Zenition 50 systems. Your local Philips representative will contact you to schedule a visit to perform this activity in your system.
Initiated
2025-09-29
Root cause
Process control
Distribution
Distribution US states of MO and TX; Belgium, Canada, Fiji, Germany, India, Malaysia, Mexico, Morocco, South Africa, South Korea, Thailand, United Kingdom, and Uzbekistan.
Recall number
Z-0241-2026
Recall status
Open, Classified
CDRH record ID
216352
FDA classified
2025-10-23
Recall event number
97682

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗