- FDA identifier
- Z-0242-2008
- Company in FDA record
- Mizuho Orthopedic Systems Inc
- FDA classification date
- Dec 13, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Failure to anchor; due to a failure in the locking mechanism, the product may not anchor properly.
Source facts
- Action
- The firm issued a Product Correction Notification letter by mail on 02/28/2007 and follow up with phone calls. The letter notified users that a required no-cost upgrade of the removal and replacement of a microprocessor that controls the logic for the floor locking mechanism will be conducted by the recalling firm.
- Initiated
- 2007-02-28
- Root cause
- Software design
- Terminated
- 2008-12-18
- Distribution
- Worldwide: USA product was distributed to 16 wholesale dealers, 212 medical facilities, 14 distributors throughout the US in the following states: AL, AZ, CA, FL, GA, HI, IL, IN, KS, KY, MA, MD, ME, MI, MO, NC, NE, NH, NJ, NM, NY, OH, OR, PA, SC, TX, VA, WA, WI, WV and WY and to the following countries Taiwan, Australia, German, Japan, Canada, Saudi Arabia, and The UK.
- Recall number
- Z-0242-2008
- Recall status
- Terminated
- CDRH record ID
- 53381
- FDA classified
- 2007-12-13
- Recall event number
- 38307