Open original FDA record ↗Recall record
- FDA identifier
- Z-0242-2022
- Company in FDA record
- GE Healthcare, LLC
- FDA classification date
- Nov 24, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The accumulated dose is incorrectly displayed in certain situations. Thisoccurs during an interventional exam using the SmartStep option if the table height is adjusted after the exam starts. In this situation, the SmartStep display will show a value that is higher (up to 100 times) than the actual dose delivered.
Source facts
- Action
- GE sent an Important Electronic Product Radiation Warning Notification letter dated October 4, 2021 to customers, The letter identified the affected product, problem and actions to be taken. Customers were informed that GE Healthcare will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services, the details of which will be included in a subsequent communication to you or through a GE Healthcare field engineer site visit. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Initiated
- 2021-06-30
- Root cause
- Radiation Control for Health and Safety Act
- Distribution
- Worldwide and US Nationwide Distribution
- Recall number
- Z-0242-2022
- Recall status
- Open, Classified
- CDRH record ID
- 190306
- FDA classified
- 2021-11-24
- Recall event number
- 89027