Recovery Compromised-during manufacturing ambient air (including oxygen) was inadvertently introduced into the culture bottles, changing their reduction/oxidation potential thus leading to a compromised recovery of some obligate anaerobes and longer detection times.
Source facts
Action
Affected Customers were notified of this recall on August 18, 2006 and August 21, 2006 via phone and instructed to destroy product remaining in stock.
Initiated
2006-08-17
Root cause
Other
Terminated
2009-12-08
Distribution
Nationwide Distribution ---- including states of AL, CA, CT, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MS, MT, NC, NE, NJ, NY, OH, PA, RI, TX, VA, WA, WI, and WV.
Recall number
Z-0243-2007
Recall status
Terminated
CDRH record ID
48238
FDA classified
2006-12-05
Recall event number
36239