Z-0243-2008: OSI, 6822 IntraOp Series Tables, Operating Room tables, Orthopedic Systems, Inc., Union City, CA. (Series

Recall record Z-0243-2008 for Mizuho Orthopedic Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0243-2008
Company in FDA record
Mizuho Orthopedic Systems Inc
FDA classification date
Dec 13, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Failure to anchor; due to a failure in the locking mechanism, the product may not anchor properly.

Source facts

Action
The firm issued a Product Correction Notification letter by mail on 02/28/2007 and follow up with phone calls. The letter notified users that a required no-cost upgrade of the removal and replacement of a microprocessor that controls the logic for the floor locking mechanism will be conducted by the recalling firm.
Initiated
2007-02-28
Root cause
Software design
Terminated
2008-12-18
Distribution
Worldwide: USA product was distributed to 16 wholesale dealers, 212 medical facilities, 14 distributors throughout the US in the following states: AL, AZ, CA, FL, GA, HI, IL, IN, KS, KY, MA, MD, ME, MI, MO, NC, NE, NH, NJ, NM, NY, OH, OR, PA, SC, TX, VA, WA, WI, WV and WY and to the following countries Taiwan, Australia, German, Japan, Canada, Saudi Arabia, and The UK.
Recall number
Z-0243-2008
Recall status
Terminated
CDRH record ID
53382
FDA classified
2007-12-13
Recall event number
38307

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗