Z-0243-2022: API 50 CH, Model 50300

Recall record Z-0243-2022 for bioMerieux, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0243-2022
Company in FDA record
bioMerieux, Inc.
FDA record creation date
Nov 17, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is potential to provide an incorrect organism identification.

Source facts

Action
Notification letters were issued to customers via FedEx Priority Overnight mail during the week of October 4, 2021. Customers are asked to do the following: -Discontinue use of and discard any remaining inventory of lot 1008679230. -For tests previously performed using API 50 CH Ref. 50300 lot# 1008679230, we are asking you to identify any possible incorrect identification test result due to a false positive reaction of test 34, analyze the related risks and to determine appropriate actions, if relevant. -Complete the Acknowledgement Form in Attachment A and return it to your local bioMrieux representative to confirm receipt of this notice. -Distribute this information to all appropriate personnel in the laboratory, retain a copy in your files. -Contact your local bioMrieux representative for product compensation. Consignees will perform destruction activities at their site.
Initiated
2021-09-30
Root cause
Material/Component Contamination
Terminated
2024-01-18
Distribution
Distribution in WI, NY, and SD
Recall number
Z-0243-2022
Recall status
Terminated
CDRH record ID
189710
FDA record created
2021-11-17
Recall event number
88844

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗