There is a risk of breakage of the torque limiting adaptor if used counterclockwise and it must not be used with a power tool due to no test data validating this use.
Source facts
Action
The recalling firm issued an Advisory Notice Information dated 2020-06-30 on 7/2/2021 via email which explained the reason for recall and that they would be contacted as soon as possible to proceed with the exchange. On 9/24/2021, a Returned Goods Authorization Form was sent to the consignees so the adapter could be exchanged.
Initiated
2021-07-02
Root cause
Nonconforming Material/Component
Terminated
2023-01-04
Distribution
Distribution was made to CO, MI, and NY. Foreign distribution was made to Canada and the United Kingdom. There was no military/government distribution.
Recall number
Z-0244-2022
Recall status
Terminated
CDRH record ID
189934
FDA record created
2021-11-17
Recall event number
88878