Skip to content

Public FDA record

Z-0244-2023: Visualase Cooled Laser Applicator System (VCLAS) 9735560 .4MM CORE FIBER 10MM TIP product. Used to necro

Recall record Z-0244-2023 for Medtronic Navigation, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0244-2023
Company in FDA record
Medtronic Navigation, Inc.
FDA classification date
Nov 17, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a defect in the outer pouch sterile seal

Source facts

Action
On 10/03/2022, the firm verbally notified affected customers and on 10/14/2022 sent an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customer informing them that Medtronic identified a breach to the sterile barrier of the outer pouch of specific lots of VCLAS 9735560 .4MM Core Fiber 10MM Tip products. Customer are instructed to: immediately identify and segregate affected products within your inventory. Medtronic will contact you to facilitate product replacement when unaffected product is available. Medtronic has determined that the inner pouches maintain sterility, accordingly, you may use segregated products ensuring the following sterile technique workflow is followed: 1. Open the affected outer pouch outside of the sterile field and remove the smaller internal pouches. 2. Per the Instruction for Use: Prior to use, examine the product packaging (in this instance the inner pouch) for damage, deterioration, and expired shelf life. If found, do not use the product. 3. Open the internal pouches using sterile technique and carefully transfer the sterile contents into the sterile field. For questions, contact Medtronic Technical Service at 1-888-826-5603 or email at rs.navtechsupport@medtronic.com
Initiated
2022-10-14
Root cause
Nonconforming Material/Component
Distribution
U.S. Nationwide distribution including in the states AZ, CA, CO, DC, FL, IL, KY, MA, MD, ME, MI, MN, MO, NH, NJ, NY, OH, PA, RI, TX, VA, WA, and WV.
Recall number
Z-0244-2023
Recall status
Open, Classified
CDRH record ID
196577
FDA classified
2022-11-17
Recall event number
91041

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗