Reason stated in the source
The firm identified circumstances in which biased results or calibration failures can be observed when using any lot of VITROS LIPA slides if they have been stored on the VITROS Chemistry Systems for more than 2 days.
Source facts
- Action
- On 10/09/06, Ortho-Clinical Diagnostics sent "IMPORTANT DISTRIBUTOR NOTIFICATION" Letters and "Confirmation of Receipt" Forms via Federal Express to all distributor accounts informing them of the "field correction", and instructing them to inform their customers who received these slides of this issue. OCD foreign affiliates were informed of this action by e-mail on 10/09/06, and instructed them to inform their customers who received the LIPA slides of this "field correction". OCD also sent "IMPORTANT PRODUCT CORRECTION NOTIFICATION" Letters and "Confirmation of Receipt" Forms to US end-user consignees that received the LIPA slides directly from OCD via US Postal Service, informing them of this "field correction" and that they may continue to use their current lot of Coating 3235 LIPA Slides until replacement product arrived providing that they followed the instructions in the letter.
- Initiated
- 2006-10-09
- Root cause
- Other
- Terminated
- 2008-05-22
- Distribution
- Worldwide distribution
- Recall number
- Z-0245-2007
- Recall status
- Terminated
- CDRH record ID
- 48782
- FDA classified
- 2006-12-05
- Recall event number
- 36512