Medicalplastic s.r.l. is conducting a voluntary field action to remove a group of Ilizarov Wire Tensioners due to a manufacturing error. An internal washer of the Ilizarov Wire Tensioner has been made in a wrong plastic material that could wear and break during the sterilization process in autoclave. Consequently this could cause malfunctioning of the tensioner.
Source facts
Action
The firm notified its sole consignee of the issue and its intent to recall by email on 06/17/2021. The recalled washers will be replaced.
Initiated
2021-06-17
Root cause
Process control
Distribution
Memphis, TN
Recall number
Z-0245-2022
Recall status
Open, Classified
CDRH record ID
189526
FDA record created
2021-11-17
Recall event number
88785
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.