Skip to content

Public FDA record

Z-0245-2023: Infection Control (IC) Lifting Strap used with the following Cirrus and Pinnacle Ceiling Lifts: Cirrus 45

Recall record Z-0245-2023 for Tollos. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0245-2023
Company in FDA record
Tollos
FDA classification date
Nov 17, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Two customers reported four incidents in which the lifting strap broke while lifting a patient. If the IC strap fails during use, the patient may fall causing injury to them or the caregiver(s).

Source facts

Action
On October 24, 2022, the firm notified affected customers via Urgent Medical Device Correction letters. The firm is updating their operator's manual and service manual with instructions to inspect the IC straps prior to use, along with photo illustrations of lifting straps demonstrating wear/damage conditions that indicate replacement must be performed. Access complete Operator s & Service Manuals on the web at www.tollos.com/service-bulletins. Tollos can be contacted by e-mail at service@tollos.com.
Initiated
2022-10-24
Root cause
Equipment maintenance
Distribution
Worldwide - US Nationwide distribution and the countries of Canada and United Arab Emirates.
Recall number
Z-0245-2023
Recall status
Open, Classified
CDRH record ID
196465
FDA classified
2022-11-17
Recall event number
91000

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗