Open original FDA record ↗Recall record
- FDA identifier
- Z-0245-2023
- Company in FDA record
- Tollos
- FDA classification date
- Nov 17, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Two customers reported four incidents in which the lifting strap broke while lifting a patient. If the IC strap fails during use, the patient may fall causing injury to them or the caregiver(s).
Source facts
- Action
- On October 24, 2022, the firm notified affected customers via Urgent Medical Device Correction letters. The firm is updating their operator's manual and service manual with instructions to inspect the IC straps prior to use, along with photo illustrations of lifting straps demonstrating wear/damage conditions that indicate replacement must be performed. Access complete Operator s & Service Manuals on the web at www.tollos.com/service-bulletins. Tollos can be contacted by e-mail at service@tollos.com.
- Initiated
- 2022-10-24
- Root cause
- Equipment maintenance
- Distribution
- Worldwide - US Nationwide distribution and the countries of Canada and United Arab Emirates.
- Recall number
- Z-0245-2023
- Recall status
- Open, Classified
- CDRH record ID
- 196465
- FDA classified
- 2022-11-17
- Recall event number
- 91000