Open original FDA record ↗Recall record
- FDA identifier
- Z-0248-2007
- Company in FDA record
- Arjo, Inc.
- FDA classification date
- Dec 6, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The sling ropes were manufactured with insufficient rope length past the knot to prevent the knot from coming loose.
Source facts
- Action
- Arjo Inc. USA sent recall letters to the direct accounts on 10/30/06, informing them that affected rope kits were incorrectly assembled, and could lead to the rope separating from the cone-shaped knob and causing the patient to fall back from the lift. The letters listed the specific quantities, order numbers and dates for the kits shipped to the customer, and requested the removal from use and return of the affected rope kits. Replacement rope kits were sent with the recall letters. Any questions were directed to Arjo''s Quality Department at 1-800-323-1245, ext. 6140.
- Initiated
- 2006-10-30
- Root cause
- Other
- Terminated
- 2008-01-18
- Distribution
- Nationwide, including Florida, Illinois, Indiana, Iowa, Kansas, Massachusetts, New York, Tennessee, Virginia and Wyoming.
- Recall number
- Z-0248-2007
- Recall status
- Terminated
- CDRH record ID
- 49156
- FDA classified
- 2006-12-06
- Recall event number
- 36680