Z-0249-2007: Sling Rope/Connector Assembly Kits for the Sara Power standing and raising aids; the kit is comprised of

Recall record Z-0249-2007 for Arjo, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0249-2007
Company in FDA record
Arjo, Inc.
FDA classification date
Dec 6, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The sling ropes were manufactured with insufficient rope length past the knot to prevent the knot from coming loose.

Source facts

Action
Arjo Inc. USA sent recall letters to the direct accounts on 10/30/06, informing them that affected rope kits were incorrectly assembled, and could lead to the rope separating from the cone-shaped knob and causing the patient to fall back from the lift. The letters listed the specific quantities, order numbers and dates for the kits shipped to the customer, and requested the removal from use and return of the affected rope kits. Replacement rope kits were sent with the recall letters. Any questions were directed to Arjo''s Quality Department at 1-800-323-1245, ext. 6140.
Initiated
2006-10-30
Root cause
Other
Terminated
2008-01-18
Distribution
Nationwide, including Florida, Illinois, Indiana, Iowa, Kansas, Massachusetts, New York, Tennessee, Virginia and Wyoming.
Recall number
Z-0249-2007
Recall status
Terminated
CDRH record ID
49371
FDA classified
2006-12-06
Recall event number
36680

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗