Z-0249-2022: Vit Kit-Freeze is intended for use in assisted reproductive procedures for vitrification and storage of h

Recall record Z-0249-2022 for Fujifilm Irvine Scientific, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0249-2022
Company in FDA record
Fujifilm Irvine Scientific, Inc.
FDA record creation date
Nov 17, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a component in kit being labeled with the incorrect Expiration Date

Source facts

Action
On September 09, 2021, the firm, Fujifilm, sent Customer notification letter via e-mail to customers (direct) and Distributors under heading "Notice: Product Labeling Information", notifying customers of the components in a kit may be labeled with an incorrect expiration date. The firm determined that the product is safe to use, however, customer to follow the expiration date on the kit label and of Certificate Of Analysis, which had shorter expiration date than the affected products expiration date. Two product types had incorrect expiration date. The customer data collection form sent to the customer of the actions to be taken by customer/end user to fill out and send the acknowledgement form if they had affected products or did not have affected products and the quantity consumed or remained in inventory and mail the acknowledgement via e-mail. The firm revised the Customer Notification letter as "Urgent: Medical Device Correction-Supplemental Information" with the Placard to customers via USPS and informed Distributors to send Notification to their customers. The customers will be asked to: - follow the expiration date on the product kit label and Certificate of Analysis - check their inventory for affected product - inform their staff of the contents of the Notification letter - place the Placard near their affected inventory - contact the firm if they prefer to return and/or replace their affected inventory. The firm provided contact information as: fisira@fujifilm.com 24/7, or by phone at 1 (949) 261-7800 The firm asked their customers in revised data collection form and added Product Information, to notify the firm on # of kits received, # of kits consumed and # of kits remaining in inventory. The Quality Assurance placed the affected lots on "cease distribution" until the firm completes an investigation.
Initiated
2021-09-08
Root cause
Software design
Distribution
Worldwide Distribution: U.S. (nationwide) to states of: AL, AK, AZ, CA, CT, DE, FL, ID, IL, IN, ND, NC, NV, NY, NJ, NE, OH, OK, OR, PA, SC, SD, MA, MS, MI, MO, MD, MN, TN, TX, UT, VA, VT, WI, WV and OUS (international) to countries of: Brazil, EU, Canada, India, South Korea, and Singapore.
Recall number
Z-0249-2022
Recall status
Open, Classified
CDRH record ID
189398
FDA record created
2021-11-17
Recall event number
88720

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗