Open original FDA record ↗Recall record
- FDA identifier
- Z-0250-2007
- Company in FDA record
- Smiths Medical ASD, Inc.
- FDA classification date
- Dec 6, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
IV Fluid Administration Set may have a blockage of the recirculating warming fluid channel within the disposable administration set, causing an alarm condition of over-temperature during priming.
Source facts
- Action
- Smiths Medical ASD notified Direct customers and distributors via letter sent by UPS on or about November 6, 2006 and the days following as necessary until all consignees are notified. Letters to direct customers will be addressed to the Risk Manager. All Smiths Medical affiliates contacted by e-mail and given instructions to send a letter to their customers and to confirm to Smiths Medical that their customers have taken appropriate action by returning the affected product. US Distributors instructed to return the affected product and to provide a customer list. Foreign distributors instructed to return the affected product and given instructions to send a letter to their customers and to confirm to Smiths Medical that their customers have taken appropriate action by returning the affected product.
- Initiated
- 2006-11-07
- Root cause
- Equipment maintenance
- Terminated
- 2009-11-30
- Distribution
- Worldwide, including USA, Japan, Canada, Ireland, UK, Finland, Netherlands, Germany, Austria, Portugal, France, Sweden, Australia, New Zealand, Russia, Italy, Singapore, Kuwait, Saudi Arabia, Czech Republic, Dubai, Spain, Norway, Denmark, and East Temor.
- Recall number
- Z-0250-2007
- Recall status
- Terminated
- CDRH record ID
- 49294
- FDA classified
- 2006-12-06
- Recall event number
- 36758