Open original FDA record ↗Recall record
- FDA identifier
- Z-0250-2013
- Company in FDA record
- Sunquest Information Systems, Inc.
- FDA classification date
- Nov 6, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Sunquest is recalling Sunquest Laboratory version 7.0 because when saving an Accession to a non-current Episode/Event, a User-Defined field can be incorrectly replaced by the current Episode/Event User-Defined data.
Source facts
- Action
- Sunquest Information Systems, Inc. sent a recall letter dated November 3, 2011, to all affected customers. Customers were informed of the problem identified and the actions to be taken. Customers were informed that a change request CR11-00174907 has been created to address this issue. This will be available as part of Sunquest Laboratory version 7.0.1. To request the software correction, customers were instructed to create a Case via the Client Support webpage at www.sunquestinfo.com/supportweb. Then choose a new case, and finally select work queue and select software request. Customers with questions were instructed to call (877) 239-6337 for US and Canadian sites and at +44 (0) 161 335 0562 for international sites. For questions regarding this recall call 520-570-2252.
- Initiated
- 2011-11-03
- Root cause
- Software design
- Terminated
- 2012-11-06
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Ireland
- Recall number
- Z-0250-2013
- Recall status
- Terminated
- CDRH record ID
- 113605
- FDA classified
- 2012-11-06
- Recall event number
- 63369