Open original FDA record ↗Recall record
- FDA identifier
- Z-0251-2020
- Company in FDA record
- Senseonics, Inc.
- FDA record creation date
- Oct 31, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Eversense Sensors have prematurely stopped functioning due to inadequate hydration of the sensor s glucose-sensing surface
Source facts
- Action
- Senseonics issued Urgent Medical Device recall letter to HCPs and Distributors as two separate pdf documents) issued on Sep 18 as UPS next day (by noon) delivery. Letter states reason for recall, health risk and action to take: If you have inserted one of these sensors, there is no reason to remove it from the patient and replace it unless it prematurely stops functioning. Replacement Alert on the Eversense app. However, if any of these sensors are in inventory awaiting insertion at your office, DO NOT INSERT THESE SENSORS. Please immediately check your inventory for the identified serial numbers and quarantine affected stock on hand. We will replace affected sensors in your inventory. Please immediately complete the attached stamped and addressed response card and check your product status. Contact your Eversense sales representative or Customer Care at 1-844 SENSE4U (844-736-7348) 24 hours/7 days-a-week if you have any questions.
- Initiated
- 2019-09-18
- Root cause
- Process control
- Terminated
- 2020-07-30
- Distribution
- Nationwide
- Recall number
- Z-0251-2020
- Recall status
- Terminated
- CDRH record ID
- 176696
- FDA record created
- 2019-10-31
- Recall event number
- 83896