Open original FDA record ↗Recall record
- FDA identifier
- Z-0252-2013
- Company in FDA record
- Diagnostica Stago, Inc.
- FDA classification date
- Nov 6, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Following a single customer complaint that incorrect INR values were occasionally printed for PT tests, even though the primary units (seconds) were correct, the firm investigated and confirmed the potential for printing incorrect secondary units when using an external printer and the customer printout mode on the STA Satellite.
Source facts
- Action
- Diagnostica Stago sent an Urgent Field Safety Notice dated June 12, 2012, to all affected customers via email return receipt requested on June 15, 2012. Customers were instructed to follow the instructions in the notification and complete the Customer Verification form and return to the Hotline by fax to 973-644-9348. For additional information, customers were instructed to contact the Hotline at 800-725-0607. For questions regarding this recall call 973-631-1200, ext 2044.
- Initiated
- 2012-06-15
- Root cause
- Software design
- Terminated
- 2013-07-19
- Distribution
- Worldwide Distribution - USA (nationwide) and Canada
- Recall number
- Z-0252-2013
- Recall status
- Terminated
- CDRH record ID
- 113273
- FDA classified
- 2012-11-06
- Recall event number
- 63270