Skip to content

Public FDA record

Z-0252-2013: STA Satellite Automated Multi-Parameter Analyzer This system is a fully automatic clinical instrument

Recall record Z-0252-2013 for Diagnostica Stago, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0252-2013
Company in FDA record
Diagnostica Stago, Inc.
FDA classification date
Nov 6, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

Following a single customer complaint that incorrect INR values were occasionally printed for PT tests, even though the primary units (seconds) were correct, the firm investigated and confirmed the potential for printing incorrect secondary units when using an external printer and the customer printout mode on the STA Satellite.

Source facts

Action
Diagnostica Stago sent an Urgent Field Safety Notice dated June 12, 2012, to all affected customers via email return receipt requested on June 15, 2012. Customers were instructed to follow the instructions in the notification and complete the Customer Verification form and return to the Hotline by fax to 973-644-9348. For additional information, customers were instructed to contact the Hotline at 800-725-0607. For questions regarding this recall call 973-631-1200, ext 2044.
Initiated
2012-06-15
Root cause
Software design
Terminated
2013-07-19
Distribution
Worldwide Distribution - USA (nationwide) and Canada
Recall number
Z-0252-2013
Recall status
Terminated
CDRH record ID
113273
FDA classified
2012-11-06
Recall event number
63270

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗