Z-0254-2018: enGen Track System with TCAutomation Software Version 4.2

Recall record Z-0254-2018 for Ortho-Clinical Diagnostics. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0254-2018
Company in FDA record
Ortho-Clinical Diagnostics
FDA record creation date
Dec 14, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A software anomaly with TCA Software V4.2 may potentially cause a delay in reporting of results due to patient samples being routed to an unintended location when samples are reintroduced back onto the enGEN track

Source facts

Action
On 26 April 2017, Ortho Clinical Diagnostic issued a notice titled "Important Product Correction Notification" and a response form (Ref. CL2017-090) to their customers via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all potentially affected enGen customers to inform them of the issue and provide them instructions on how to monitor and correct the issue if it has occurred. Foreign affiliates were informed by email on 26 April 2017 of the issue. Potentially affected customers are those that have enGen Systems with: 1) TCA software V4.2 and configured with Rack Exit Module(s) and/or a Rack Entry & Exit (Combined) Module(s), but not configured with any ES Flex modules, and/or 2) TCA software V4.2 and a Stago analyzer with the Stago STA-R interface. **Required Actions- 1) When checking samples in the Check drawer, do the following based on your enGen System configuration: -For enGen Systems configured with Rack Entry & Exit (Combined) Module or Rack Exit Module without ES Flex, you may observe an increase in Sample IDs (SID) with General Error Duplicate SID in router messages in the Check Window. Remove the samples that are falsely identified as Duplicate SIDs from the Check drawer and reload the sample on the track for further processing if needed. NOTE: Refer to the Question and Answer section on page 3 for how to identify which samples are falsely identified as Duplicate SIDs. - For enGen Systems configured with a Stago analyzer, you may observe and increase in SIDs with "Sample not defined in router" messages for samples returning from the analyzer. Perform all post-analytical processing steps for associated samples, such as recapping, sorting and storing, off-track (manually). 2) Complete and return the attached Confirmation of Receipt form by May 9, 2017. Customers with questions may contact the Technical Solutions Center at 1-800-421-3311, 24 hours per day 7 days per week
Initiated
2017-04-26
Root cause
Software design (manufacturing process)
Terminated
2021-08-23
Distribution
Domestically - GA, IL, MO, NC & NY; Internationally - Canada, England, France, Italy & Spain
Recall number
Z-0254-2018
Recall status
Terminated
CDRH record ID
155157
FDA record created
2017-12-14
Recall event number
77172

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗