Mislabeling: Reporting terminology in the Syphilis IgG APF CD is not consistent with the distributed Instructions for Use. (Non-Reactive and Reactive rather than Negative and Positive)
Source facts
Action
The recalling firm notified consignees by phone call on 9/25/2007, followed up with a fax notification. The firm plans to monitor the consignees that do not respond by tracking on a spreadsheet and following up with another phone call. Consignees are instructed to destroy the APF CD which will be replaced at a later date.
Initiated
2007-09-25
Root cause
Labeling Change Control
Terminated
2008-04-16
Distribution
Nationwide
Recall number
Z-0256-2008
Recall status
Terminated
CDRH record ID
65277
FDA classified
2007-11-09
Recall event number
45332