Sample problems; Five issues identified related to software utilized by the ACCELERATOR APS System Input/Output Module, to include: Under three specific conditions, the post-aspiration radio-frequency identification (RF ID) read of the tube carrier is not performed and the APS work cell may not appropriately generate Sample Presentation/Sample Queue errors with APS software version 1.1.1 or earlie
Source facts
Action
Firm notified consignee via letter on 11/10/07. Consignee notified of software issues and informed they would be contacted by an Abbott representative to schedule installation of software upgrade.
Initiated
2007-10-10
Root cause
Software design
Terminated
2008-03-10
Distribution
Nationwide: including state of UT.
Recall number
Z-0257-2008
Recall status
Terminated
CDRH record ID
65475
FDA classified
2007-11-24
Recall event number
45429