Open original FDA record ↗Recall record
- FDA identifier
- Z-0257-2013
- Company in FDA record
- Kimberly-Clark Corporation
- FDA classification date
- Nov 7, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Package seam of some units may not remain properly sealed to ensure the contents might not be sterile as labeled.
Source facts
- Action
- Kimberly Clark sent an Urgent Voluntary Product Recall letter dated July 5, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to determine if they have any affected product in their inventory. Any affected product should be quarantined and distribution discontinued. Distributors were also asked to conduct a sub-recall for any affected product further distributed. Customers were asked to complete the attached Distributor Product Recall Response Form and fax to 1-920-225-3522. Upon completion of the response form customers were asked to destroy any affected product. For questions regarding this recall call 770-587-7835.
- Initiated
- 2012-07-05
- Root cause
- Packaging process control
- Terminated
- 2013-06-13
- Distribution
- Nationwide Distribution
- Recall number
- Z-0257-2013
- Recall status
- Terminated
- CDRH record ID
- 114310
- FDA classified
- 2012-11-07
- Recall event number
- 63598