Z-0257-2020: CardioLab/ComboLab Recording Systems

Recall record Z-0257-2020 for GE Healthcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0257-2020
Company in FDA record
GE Healthcare, LLC
FDA record creation date
Oct 31, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for failure of the patient leakage current test. There is a potential that if another device with electrical connectivity (e.g., ablation device, ECG monitor) fails, then the CLab II Plus Amplifier will not prevent an electric current from completing a circuit, which could cause an electrical shock to a patient.

Source facts

Action
GE Healthcare notified customers of the recall on about 09/16/2019, via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were informed that certain CLab II Plus Amplifiers used with CardioLab/ComboLab systems have failed the patient leakage current test. There is a potential that if another device with electrical connectivity (e.g., ablation device, ECG monitor) fails, then the CLab II Plus Amplifier will not prevent an electric current from completing a circuit, which could cause an electrical shock to a patient. Instructions included that customers could continue to use the system by following the below instructions: 1. Test all equipment used in conjunction with CardioLab/ComboLab per manufacturer instructions on electrical safety to ensure that all equipment is in good operating condition and performs per manufacturer specifications. Note: Due to the potential safety risks of other devices used in conjunction with CLab II Plus Amplifiers in Electrophysiology Procedures, all equipment that can electrically contact the patient during a procedure should be checked against each devices safety conformances in manufacturer specifications. 2. If you suspect another connected device is or may be malfunctioning, discontinue use of the CLab II Plus Amplifier until all nonconforming equipment is corrected. If nonconforming equipment is used in conjunction with the CLab II Plus Amplifier, the CLab II Plus Amplifier will not provide necessary mitigation to prevent patient shock. Customers were also informed that GE Healthcare will correct all affected products and a GE Healthcare representative will contact customers to arrange for the correction. They were asked to complete and return the "MEDICAL DEVICE NOTIFICATION ACKNOWLEDGEMENT" form.
Initiated
2019-09-16
Root cause
Device Design
Terminated
2020-12-16
Distribution
Nationwide distribution to CA, CT, FL, IN, MD, MI , MN, MO, NC, NE, NJ, OH, PA, TX, WI. International distribution to Canada, France, India, Japan, Korea (Republic Of), Mexico, Spain, Taiwan.
Recall number
Z-0257-2020
Recall status
Terminated
CDRH record ID
176715
FDA record created
2019-10-31
Recall event number
83899

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗