Z-0258-2020: The Infinity Acute Care System (IACS) Monitoring Solution with the Standalone Infinity M540 patient monit

Recall record Z-0258-2020 for Draegar Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0258-2020
Company in FDA record
Draegar Medical Systems, Inc.
FDA classification date
Oct 31, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Cybersecurity vulnerabilities may cause device to reboot, lose alarm functionality, and/or lose communication with cockpit and/or the Infinity Network.

Source facts

Action
On September 13, 2019, the firm distributed Urgent Medical Device Recall letters to customers. Customers were informed of the cybersecurity vulnerabilities, which can include Distributed Denial of Service (DDOS) (packet storm), Spoofing, and Tampering. To mitigate the cybersecurity concerns, Draeger will be releasing software version VG4.2 for both the Cockpit and the M540, which will correct these cybersecurity vulnerabilities. The software is expected to be released for distribution in December 2019. Upgrades of the IACS systems will commence in January 2020. While the firm is in the process of updating the software, customers are asked to limit access to the Infinity Network by following these security recommendations: 1. Physical security of the patient monitors is recommended and is the responsibility of the operating organization. 2. Physical security of the telecommunications closet is recommended and is the responsibility of the operating organization. 3. Draeger recommends that operating organizations restrict physical access to unused Ethernet ports on the IACS. 4. Draeger recommends that operating organizations restrict physical access to unused USB and serial ports on the IACS. 5. Draeger relies on the medical device isolation mechanism of the VLANs and the proper configuration, implementation, and use of the operating organization's security measures to prevent the introduction of malware onto the Infinity Network. Your local Draeger Service Representative will contact you to schedule an appointment to upgrade your system(s) software free of charge once the new software version is released for distribution. If you have any questions regarding this letter, please contact Michael Kelhart between the hours of 8:00 AM - 4:30 PM EST at 1-800-437-2437 (press 1 at the prompt, then 32349).
Initiated
2019-09-13
Root cause
Software design
Distribution
US Nationwide distribution.
Recall number
Z-0258-2020
Recall status
Open, Classified
CDRH record ID
176681
FDA classified
2019-10-31
Recall event number
83889

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗