Open original FDA record ↗Recall record
- FDA identifier
- Z-0258-2024
- Company in FDA record
- Encore Medical, LP
- FDA classification date
- Nov 7, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Knee punch handle tip is breaking during surgery due to the lack of heat treatment, which may lead to a disruption in surgery.
Source facts
- Action
- On 9/29/23, recall notices were emailed to customers who were asked to do the following: 1) Knee punch handles should be visually inspected prior to use for any damage. Upon completion of surgery, visually inspect and verify there is no breakage. The attached quality bulletin provides guidance on an acceptable and unacceptable device. 2) Contact your customer service rep at customerservice@enovis.com to report failures or to request replacements. 3) Complete and return the response form that asks customers to check all related devices in inventory and report the number of related devices in inventory. Questions can be directed to livia.kuruvila@enovis.com
- Initiated
- 2023-09-29
- Root cause
- Process change control
- Distribution
- US: VA, TN, MN, TX, IL, IN, KY, OH, SC, MS, RI, ME, NJ, NY, MD, CA, KS, LA, PR, AL, FL, GA, AZ, WA
- Recall number
- Z-0258-2024
- Recall status
- Open, Classified
- CDRH record ID
- 203678
- FDA classified
- 2023-11-07
- Recall event number
- 93214