Open original FDA record ↗Recall record
- FDA identifier
- Z-0258-2025
- Company in FDA record
- DRG International, Inc.
- FDA classification date
- Oct 29, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
The High and Low Control acceptance ranges in the Certificate of Analysis (CofA) did not match the acceptance range information on the High and Low Control Component labels. Due to the defect, the device failed to provide results, which may result in a delay in obtaining test results and a remote chance of an adverse health risk.
Source facts
- Action
- DRG issued Urgent Field Safety Notification on 9/17/24 via email. Letter states reason for recall, health risk and action to take: Please check your inventory for the respective lot and ensure that the proper acceptance ranges provided below are used to validate the assay. Immediately reply to email: qa@drg-international.com to confirm this notice and your compliance.
- Initiated
- 2024-09-13
- Root cause
- Process change control
- Distribution
- AZ, CA, NH Foreign: Austria Belgium Georgia Germany Greece India Irak Italy Netherlands Poland Romania Russia Singapore Spain Tunesia Vietnam
- Recall number
- Z-0258-2025
- Recall status
- Open, Classified
- CDRH record ID
- 210237
- FDA classified
- 2024-10-29
- Recall event number
- 95429