Skip to content

Public FDA record

Z-0258-2025: 17-OH Progesterone ELISA 96 well kit- An enzyme immunoassay for the quantitative in vitro diagnostic meas

Recall record Z-0258-2025 for DRG International, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0258-2025
Company in FDA record
DRG International, Inc.
FDA classification date
Oct 29, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

The High and Low Control acceptance ranges in the Certificate of Analysis (CofA) did not match the acceptance range information on the High and Low Control Component labels. Due to the defect, the device failed to provide results, which may result in a delay in obtaining test results and a remote chance of an adverse health risk.

Source facts

Action
DRG issued Urgent Field Safety Notification on 9/17/24 via email. Letter states reason for recall, health risk and action to take: Please check your inventory for the respective lot and ensure that the proper acceptance ranges provided below are used to validate the assay. Immediately reply to email: qa@drg-international.com to confirm this notice and your compliance.
Initiated
2024-09-13
Root cause
Process change control
Distribution
AZ, CA, NH Foreign: Austria Belgium Georgia Germany Greece India Irak Italy Netherlands Poland Romania Russia Singapore Spain Tunesia Vietnam
Recall number
Z-0258-2025
Recall status
Open, Classified
CDRH record ID
210237
FDA classified
2024-10-29
Recall event number
95429

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗