An error code (Code 13) is displayed on some of the consoles when a Medtronic Stylus or EHS handpiece is connected to the IPC indicating that the handpiece has not been recognized by the console.
Source facts
Action
A Product Correction Notification was issued to each consignee on June 15, 2009. Medtronic sales representatives will visit each facility to advise consignees of the issue and to install a software upgrade that will correct the handpiece recognition issue on the console(s).
Initiated
2009-07-17
Root cause
Software design
Terminated
2010-01-20
Distribution
Worldwide Distribution
Recall number
Z-0259-2010
Recall status
Terminated
CDRH record ID
85967
FDA classified
2009-11-13
Recall event number
53603