Skip to content

Public FDA record

Z-0259-2024: Monitron II Waveform Analyzer, REF: S00015-1, used with ventilators.

Recall record Z-0259-2024 for Percussionaire Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0259-2024
Company in FDA record
Percussionaire Corporation
FDA classification date
Nov 7, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Note this recall occurred in 2020 and 2021. Reports of screen freeze on visual displays of waveform analyzers that are paired with ventilators. Subsequently, waveform analyzers were serviced to install PAL microprocessor chips and firmware.

Source facts

Action
Waveform analyzers were serviced from September 2020 through September 2021.
Initiated
2020-09-14
Root cause
Device Design
Distribution
US: CA, ND, MS, MD
Recall number
Z-0259-2024
Recall status
Completed
CDRH record ID
203628
FDA classified
2023-11-07
Recall event number
93192

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗