Open original FDA record ↗Recall record
- FDA identifier
- Z-0259-2025
- Company in FDA record
- Princeton Biomeditech Corp
- FDA classification date
- Oct 30, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
This recall was initiated due to customer complaints of false negative of hCG test results when running patient urine samples. Use of the above listed product lot may result in false negative results. An end user reported that the OSOM Ultra hCG Combo Test, lot 563L13, produced false negative test results from 2 patient urine samples as confirmed by quantitative hCG and ultrasound. During further root cause testing and investigation, an additional 6 devices were found to produce false negative test results from hCG positive samples.
Source facts
- Action
- PBM notified the Distributor Sekisui Diagnostics on 8/23/24 via email. The sent an "URGENT PRODUCT RECALL NOTICE" dated 8/30/24 to customers. Sekisui Diagnostics issued a recall notice on 9/6/24 to its customers. PBM Letter states reason for recall, health risk and action to take: Stop distribution or use of OSOM Ultra hCG Combo Test lot 563L13. " If you have further distributed this product, please identify your customers, and notify them at once of this product recall. This recall should be carried out to the user level. Your notification to your customers may be enhanced by including a copy of this recall notification letter. " Quarantine and return any kits from the listed lots, (1) currently in your inventory, and (2) returned from your customers. " Indicate the quantity returned on the attached CONFIRMATION OF NOTIFICATION form, and follow the instructions to complete, even if you do not have any affected product remaining in your inventory. Complete, sign and return the CONFIRMATION OF NOTIFICATION form to the email provided by the date indicated on the form. Please note replacement product equivalent to the quantity of products you returned from the affected lot will be processed once the CONFIRMATION OF NOTIFICATION form has been received. If you have any questions, call PBM Technical Services at Tel 1-732-274-1000 or E-mail .
- Initiated
- 2024-08-23
- Root cause
- Process change control
- Distribution
- US Distribution to MO only.
- Recall number
- Z-0259-2025
- Recall status
- Open, Classified
- CDRH record ID
- 210290
- FDA classified
- 2024-10-30
- Recall event number
- 95439