Skip to content

Public FDA record

Z-0260-2020: Dario Blood Glucose Monitoring System, Model No. 1015-05, running Android software versions 4.3.0-4.3.2

Recall record Z-0260-2020 for LABSTYLE INNOVATIONS. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0260-2020
Company in FDA record
LABSTYLE INNOVATIONS
FDA record creation date
Oct 31, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Dario-Blood Glucose Tracker Android App versions 4.3.0-4.3.2 may experience duplicate logging of a blood glucose level reading.

Source facts

Action
The firm released Android App version 4.3.3 on August 29, 2019. Users were informed about the issue via push notification on 8/30/19, asking to update to the new version. An additional push notification was sent on 8/31/19, along with a Field Safety Notice letter sent to their "inbox". On 9/9/19, remaining potential relevant users received a third notification in which they were asked to upgrade their version through in-app and push notifications. Customers were instructed to do the following: Please update the latest Dario - Blood Glucose Tracker Android App version as follows o Open the Google Play store o Search for Dario Blood Glucose Tracker App o Press Uninstall button o Press OK to confirm the question "Do you want to uninstall this app?" o Once completed, press Install and verify you installed 4.3.3 version (or above) by pressing the "About this App" link. Customers with questions should contact the firm's Customer Service Center at 1-800-895-5921, Monday Friday, 9AM 5PM Eastern.
Initiated
2019-08-29
Root cause
Software change control
Terminated
2020-05-05
Distribution
Nationwide domestic distribution. Foreign distribution to Australia, Canada, UK, and Italy.
Recall number
Z-0260-2020
Recall status
Terminated
CDRH record ID
176305
FDA record created
2019-10-31
Recall event number
83761

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗