The patient label stickers may have the wrong expiration date recorded. The contract manufacturer entered the manufacturing date (2024-04-01) as the expiration date. The actual expiration date of the product is 2027-04-01. This is reflected on the carton label and sterile pouch label correctly.
Source facts
Action
TrackX Technology issued an Urgent: Medical Device Field Correction notice to its consignees via hand delivery by their sales representatives on 10/01/2024. The notice explained the issue and requested that the notice be provided to all parties to whom the product was distributed. The sales representatives exchanged the affected product for correctly labeled product.
Initiated
2024-09-03
Root cause
Error in labeling
Distribution
NC, VA, FL, PA, NY
Recall number
Z-0260-2025
Recall status
Open, Classified
CDRH record ID
210214
FDA classified
2024-10-30
Recall event number
95424