- FDA identifier
- Z-0261-2008
- Company in FDA record
- Arjo, Inc.
- FDA classification date
- Dec 1, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Male patient genitals may be pinched between the seat and the frame of the patient lift chair when the Care Raiser seat is returned to its lowest position.
Source facts
- Action
- Arjo sent Customer Field Correction Notification letters dated 10/15/07 to the end user accounts who received the affected chairs, advising them of the potential for the pinching of male patient genitals between moving parts of the care raiser function of the Carendo Hygiene Chair. They were instructed to review the enclosed revised Operating and Product Care Instructions manual, highlighting the warnings on pages 18 and 20; ensure all users receive refresher training in the chair's proper use and have access to the revised manual; and complete and return the enclosed customer response form, acknowledging the receipt and understanding of the letter. Any questions were directed to the Arjo Quality Department at 800-323-1245, ext. 6140 or 6165.
- Initiated
- 2007-10-15
- Root cause
- Device Design
- Terminated
- 2009-07-27
- Distribution
- Nationwide
- Recall number
- Z-0261-2008
- Recall status
- Terminated
- CDRH record ID
- 65381
- FDA classified
- 2007-12-01
- Recall event number
- 45402