Open original FDA record ↗Recall record
- FDA identifier
- Z-0261-2012
- Company in FDA record
- SCC Soft Computer
- FDA classification date
- Nov 21, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
On 08/08/2011 SCC Soft Computer, Clearwater, FL initiated a correction on following SoftLab used with: SA BASE 3.1.4.x, 3.1.6.x and SA LabMic 4.0.0.x, 4.0.1.x, 4.0.2.x, 4.0.3.x, 4.0.4.x. A client reported a flagging error where the result was flagged as critical in the Instrument Menu but the printed report did not show the corresponding flag. All affected clients were notified of the issue on 0
Source facts
- Action
- SCC Soft Computer notified all affected customers on August 8, 2011, of the recall via the SCC Soft Computer proprietary Task management system to manage client communication. Clients use this tool to log in requests for services and receive and provide updates to any issue. Each client user has a unique password and every action is date and time-stamped. Each Task has a unique identifier and a full audit trail. SCC recommended to the customer until Live on the correction avoid making edits in the instrument menu. For further questions please call (727) 789-0100.
- Initiated
- 2011-08-08
- Root cause
- Software design
- Terminated
- 2013-11-22
- Distribution
- Worldwide Distribution - USA ( nationwide) and Canada.
- Recall number
- Z-0261-2012
- Recall status
- Terminated
- CDRH record ID
- 105257
- FDA classified
- 2011-11-21
- Recall event number
- 60405