Skip to content

Public FDA record

Z-0261-2022: Quantum 2000 Electrosurgical Generators Cooper Surgical Part Number: 909075

Recall record Z-0261-2022 for CooperSurgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0261-2022
Company in FDA record
CooperSurgical, Inc.
FDA record creation date
Nov 18, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Coagulate function may not operate with the use of the Hand Switch, failure of the device to coagulate tissue may lead to a prolonged procedure requiring medical intervention to control any intraoperative bleeding and complete the case.

Source facts

Action
CooperSurgical issued Urgent Medical Device Recall to consignees via certified mail with confirmed delivery receipt dated 9/29/21. Letter states reason for recall, health risk and action to take: This Recall affects 2955 units manufactured since product launch in December 2009 and a corrective action has been initiated to prevent future recurrence. Please discontinue use of this product, quarantine the product, and complete the attached Acknowledgement and Receipt Form to arrange for repair of your device at CooperSurgical s expense, including shipping, and return of the product to you. Please feel free to reach us at 203-601-5200 ext. 3300. Please complete this form and return it via email: recall@coopersurgical.com or fax to 203.601.9870 ATTN: Product Surveillance. CooperSurgical will arrange for a product repair after this form has been received. Firm issued additional letter for the product LP-10-220 as it was not included in the initial recall letter. The letter dated 10/18/21 distributed to international accounts only.
Initiated
2021-09-29
Root cause
Component design/selection
Distribution
US Nationwide Distribution Foreign: ARMENIA, BOLIVIA, CAMBODIA, CANADA, CHILE, CHINA, COSTA RICA, DOMINICAN REPUBLIC, ECUADOR, ENGLAND, GERMANY, HONG KONG, INDIA, INDONESIA, IRAN, ITALY, JORDAN, KENYA, MALAYSIA, MAURITIUS, MEXICO, NEW ZEALAND, NICARAGUA, NIGERIA, PHILIPPINES, PUERTO RICO, QATAR, RWANDA, SINGAPORE, SOUTH AFRICA, SRI LANKA, TAIWAN, TANZANIA, THAILAND, TRINIDAD, TURKEY, UGANDA, UNITED ARAB EMIRATES, VIETNAM
Recall number
Z-0261-2022
Recall status
Open, Classified
CDRH record ID
189194
FDA record created
2021-11-18
Recall event number
88648

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗