Beckman Coulter has confirmed that occasionally a cuvette may be skipped during the SYNCHRON Lipase Wash (LIWA) procedure. If this occurs, there is the potential for carryover into the SYNCHRON Lipase (LIPA) assay, resulting ina substantial positive bias affecting Lipase (LIPA) results.
Source facts
Action
The week of 10/25/2004, a product corrective action (PCA) letter was mailed vis U.S. mail to all customers shipped LIPA reagent for the 24 months preceding the discovery of the issue. Customers instructed to repeat any sample that produces a lipase value of > 170 U/L that cannot be confirmed by patient history.
Initiated
2004-10-25
Root cause
Other
Terminated
2005-06-17
Distribution
US and Canada
Recall number
Z-0262-5
Recall status
Terminated
CDRH record ID
35734
FDA classified
2004-11-30
Recall event number
30362