Z-0263-2025: 1) Flex Extenders, 5.2" x 15mm OD x 15mm ID REF S6411P 2) Flex Extender, 5.2" x 15mm OD x 22/15mm ID REF

Recall record Z-0263-2025 for Westmed, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0263-2025
Company in FDA record
Westmed, Inc.
FDA classification date
Oct 30, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to connector within the circuit flex extender not meeting specifications resulting in a low connection force and as a result, the connector may leak or disconnect and may cause life threating conditions or death.

Source facts

Action
On 09/27/2024, A written customer recall notification in the form of an URGENT: Customer Recall Letter was provided to consignees that were sent the affected lots of Westmed Flex Extenders. To address the potential patient safety risk, AirLife statement to distributors and end-users of Westmed Flex Extenders is: Customer immediate actions: 1. Review the list of affected products. Examine their inventory for the mentioned lot(s). No other lots are affected. 2. Complete and return the attached Response Form via email to: productquality@myairlife.com as soon as possible. This will enable that the firm documents the amount of product customers have on hand for return and/or destruction. It will also allow for documentation of customers' receipt of this letter. 3. In addition, if product was further distributed, please identify customers/consignees, and notify them of this product removal. Further notification may be enhanced by including a copy of this removal notification letter. If any assistance contacting your customers is needed, let Recalling Firm know and they will communicate this removal to them on your behalf. 4. Once customers return and/or confirm destruction of the affected product(s), new replacement product will be sent. If replacement products are needed to be sent urgently, call directly at 1 800 433 2797 and Recall Firm will make every effort to accommodate your needs. 5. Make sure that all affected personnel in your organization are informed of this removal notice. For any questions regarding this field action, call AirLife at 1 800 433 2797, or e mail productquality@myairlife.com.
Initiated
2024-09-27
Root cause
Under Investigation by firm
Distribution
U.S.: AK, CA, CO, DC, FL, GA, IL, IN, KY, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OR, PA, SC, TX, VA, VT, WA, and WV O.U.S.: N/A
Recall number
Z-0263-2025
Recall status
Open, Classified
CDRH record ID
210326
FDA classified
2024-10-30
Recall event number
95453

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗