Sterility (package integrity) compromised: package sealing problem leaving packaging being partially open on one of the four sealed Tyvek pouch edges.
Source facts
Action
B. Braun sent an Urgent Medical Device Removal letter (10/10/07) by express mail to each of the hospital accounts that received the recalled lot of VenaTech LP. B. Braun sales represenatatives will also contact each account directly to inquire if any inventory of the suspect lot remains and to provide guidance for return and replacement of any existing inventory.
Initiated
2007-10-30
Root cause
Packaging process control
Terminated
2008-03-02
Distribution
Nationwide: States of AL, GA, PA, FL, NE, WI, IL, LA, TX, AZ, MN, MA, CA, NY, OH, IN, and MI.
Recall number
Z-0264-2008
Recall status
Terminated
CDRH record ID
65749
FDA classified
2007-11-24
Recall event number
45539