Z-0264-2018: ETEST OXACILLIN OX 256 US F100 In vitro diagnostic, a quantitative technique for determining the anti

Recall record Z-0264-2018 for BioMerieux SA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0264-2018
Company in FDA record
BioMerieux SA
FDA classification date
Dec 19, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

QC failures (MIC out of range high) for S.aureus ATCC 29213 strain on ETEST OX 256 (OXACILLIN) Foam packaging (Ref. 520558, 520518) reported. The product behavior could lead to False Resistant results.

Source facts

Action
BioMerieux sent an Urgent Field Safety Notice on April 30, 2017. Short-term Action: Subsidiaries and distributors are required to discard all ETEST OX 256 (OXACILLIN) Foam packaging (Ref. 520558, 520518) under (PSS#3320) remaining in their inventory. In order to avoid any backorders on the field during the transition to the SPB configuration that is conformed to the expected specifications, Subsidiaries and Distributors were instructed to distribute the Urgent Field Safety Notice (customer letter) to inform Medical customers and Veterinary customers that have received and will receive the product. The Urgent Field Safety Notice informed customers of the non-conformity along with instructions to continue the use of the impacted products with recommendations described in customer letter, and in addition, for the Medical Customers, to discuss any concerns they may have regarding previously reported Susceptible results on the impacted batches Listed Table 1 with their Laboratory Medical Director to determine the appropriate course of action. Long-term Action: A CAPA (PR#1101306) has been initiated in order to identify the root-cause and prevent the issue to recur. They were instructed to complete and return the attached Acknowledgement Form by Fax to confirm receipt of this notice. For further questions, please call (314) 731-8694.
Initiated
2017-04-13
Root cause
Under Investigation by firm
Terminated
2020-08-18
Distribution
Worldwide Distribution - USA (nationwide) and to the countries of : Algeria, Austria, Australia, Belarus, Bosnia-Herz., Canada, Croatia, Czech Republic, Egypt, Estonia, France, Georgia, Germany, Greece, Guam, Hong Kong, Hungary, India, Israel, Italy, Japan, Kuwait, Lebanon, Malta, Mexico, Netherlands, Nigeria, Peru, Philippines, Qatar, Russian Fed., Saudi Arabia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom and Utd.Arab Emir.
Recall number
Z-0264-2018
Recall status
Terminated
CDRH record ID
160120
FDA classified
2017-12-19
Recall event number
78688

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗