Z-0264-2025: Olympus Electrosurgical Generator, Model Number ESG-410 (SEP) Version WA91327U

Recall record Z-0264-2025 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0264-2025
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Oct 30, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There was an increase in complaints indicating an occurrence of system error E1226 "Could not write to Audio Buffer" during procedures using the ESG-410 (SEP).

Source facts

Action
An URGENT: MEDICAL DEVICE CORRECTIVE ACTION notification letter dated 10/1/24 was sent to customers. Actions Required: Our records indicate that your facility has received one or more affected ESG-410 (SEP) units. An Olympus representative will reach out to you to coordinate a mutually convenient time to complete a software update on your affected ESG-410 (SEP) unit(s). You may continue to use your ESG-410 (SEP) generator until the software update is completed. Additionally, Olympus requires you to take the following actions: 1. Carefully read the content of this notification. 2. Olympus requests that you acknowledge receipt of this letter. Access the Olympus recall portal to indicate that you have received this notification. " Go to https://olympusamerica.com/recall Enter the recall number "0457 " Complete the form as instructed. 3. Olympus s record of the completed software upgrade to your ESG-410 (SEP) generator(s) will serve as the acknowledgment of this field corrective action for your facility 4. If you have further distributed this product, identify your customers, and forward them this notification. Olympus requests that you report any complaints, including the ESG-410 generator system error E1226 "Could not write to Audio Buffer , or any associated injuries to the Technical Assistance Center (TAC) at 1- 800-848-9024, option 1, and the FDA. If you require additional information, please do not hesitate to contact me directly by phone at (647) 999-3203 or by e-mail at Cynthia.Ow@Olympus.com.
Initiated
2024-09-04
Root cause
Software design
Distribution
US Nationwide.
Recall number
Z-0264-2025
Recall status
Open, Classified
CDRH record ID
210352
FDA classified
2024-10-30
Recall event number
95477

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗