Z-0264-2027: Generic Name: restor3d MTP Implant Generic Name: MTP Implant Model/Catalog Number: 1103-161620 Software V

Recall record Z-0264-2027 for restor3d Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0264-2027
Company in FDA record
restor3d Inc.
FDA classification date
Oct 22, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Implants were found in the field with orange sterilzation indicators which means implants were not exposed to sterilization.

Source facts

Action
On 09/21/2026, an "URGENT: MEDICAL DEVICE VOLUNTARY RECALL" Letter was emailed to customers informing them that, Restor3d has identified an error in the sterilization of a lot of large MTP implants (part number 1103-161620) were found in the field with orange sterilization indicators, rather than red, when orange indicates not exposed to sterilization and red indicates exposed to sterilization. Root cause investigation to date indicates that the lot was never sterilized. Customers are instructed to: Do not implant or otherwise use any affected units identified above. Quarantine all affected, unused units identified above and remove them from implant sets, consignment inventory, inventory shelves, and any other storage locations. Complete the attached Field Action Response Form (Attachment A) and return it to restor3d within 10 business days, whether or not affected product is in your possession, so that restor3d can confirm effective distribution of this notice. Package the affected MTP Implants (1103-161620; A02979) to return to restor3d. restor3d has provided a prepaid FedEx shipping label by email for shipment back to 4001 NC-54 Hwy, Suite 2160 Durham, NC 27709. Promptly report to restor3d any patient adverse event, suspected device-related complication, revision surgery, or other clinical concern potentially associated with the affected product. For questions or further assistance - contact Restor3d can be reached at 984-888-0593 or at complaints@restor3d.com.
Initiated
2026-09-04
Root cause
Process change control
Distribution
U.S. Nationwide distribution in the states of FL, IL, NC, PA, SC, and TX.
Recall number
Z-0264-2027
Recall status
Open, Classified
CDRH record ID
222945
FDA classified
2026-10-22
Recall event number
99940

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗