Open original FDA record ↗Recall record
- FDA identifier
- Z-0265-2020
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- Nov 21, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
SOMATOM go.Top (Models #11061640 & 11061648) with software syngo CT VA20A_SP2 and active Guide&GO license No dose documentation and no Dose Alert for the special mode i- Sequence during interventional procedures
Source facts
- Action
- Siemens issued a Customer Safety Advisory Notice to customers via CT072/19 on 9/24/2019. The letter identifies the problem, health risk and action to take: To avoid this issue, please refrain from using the i-Sequence scan mode in combination with the 3x3 mm collimation until this issue has been solved. Please use the i-Sequence with the 3x5 mm collimation instead, as here the correct dose information will be shown. This issue will be resolved with an improved software version syngo CT VA20A_SP2b. The software update will be released, free of charge, with the update CT073/19/S for all affected systems. The software update release is planned for Q3/2019. All future service packs will contain the update as well. Questions: contact service organization at 1-800-888-7436.
- Initiated
- 2019-09-24
- Root cause
- Software design
- Terminated
- 2024-07-08
- Distribution
- US in the states of: GA, IN, MO, MS, NJ, OH, PA
- Recall number
- Z-0265-2020
- Recall status
- Terminated
- CDRH record ID
- 176670
- FDA record created
- 2019-11-21
- Recall event number
- 83883