Z-0265-2025: Access Thyroglobulin Antibody II, REF A32898, containing Instructions for Use, IFU REF B14994, B63978, an

Recall record Z-0265-2025 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0265-2025
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Oct 30, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

IFUs for the device incorrectly states the biotin interference was remediated. The issue could result in potentially false-low results.

Source facts

Action
The recalling firm issued letters dated August 28, 2024, which were issued 9/5/2024 via email and first class mail. The letter describes the issue, impact, and actions that should be taken. The consignee is to retain a copy of the letter as it serves as current labeling regarding biotin interference with the Access Thyroglobulin Antibody II assay. Appendix A of the letter provides the incorrect and correct sections of the IFUs. The recalling firm recommends sharing the content of the letter with their laboratory and/or medical director to assess the need for retesting. The letter informs the consignee that Beckman Coulter will update the IFUs to contain accurate biotin interference claims for the assay. Notifications will be sent when the updated IFU is available for download from the Beckman Coulter website. The letter also reports that IFU part number A38168 has been retired and will not be updated. If the affected product has been forwarded to another laboratory, the consignee is requested to provide them a copy of the letter. The consignee is to respond within 10 days electronically via email or manually by completing and returning the enclosed response form.
Initiated
2024-09-05
Root cause
Error in labeling
Distribution
Distribution was made nationwide to the U.S., including Puerto Rico. There was government distribution but no military distribution. Foreign distribution was made to Mexico.
Recall number
Z-0265-2025
Recall status
Open, Classified
CDRH record ID
210145
FDA classified
2024-10-30
Recall event number
95407

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗