Open original FDA record ↗Recall record
- FDA identifier
- Z-0265-2027
- Company in FDA record
- Medtronic MiniMed, Inc.
- FDA classification date
- Oct 22, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A software defect in which existing users who had previously configured and enabled the long-acting dose reminder feature in version 1.2.0 or earlier experienced failure of the long-acting dose reminder after updating their app to version 1.2.1. The long-acting dose reminder is an optional feature that, when enabled, triggers a notification at the scheduled time to remind users to take their long-acting insulin dose. The issue was caused by a type cast exception during database migration that occurs upon app update to version 1.2.1 causing a silent failure that prevented the creation of new alerts. Failure to receive the long-acting dose reminder may result in missed or delayed insulin doses, potentially leading to hyperglycemia. Conversely, if users are uncertain whether they took their dose due to the missing reminder, inadvertent duplicate dosing may occur, potentially resulting in hypoglycemia.
Source facts
- Action
- On 09/14/2026, customers were contacted via SMS, emailed or via an "URGENT MEDICAL DEVICE CORRECTION" Letter informing them that, an issue was identified with the MiniMed Go app in which the long-acting dose reminders were not being delivered after the app was updated to version 1.2.1. Actions required: Users of MiniMed Go app software versions 1.2.1 should update to version 1.2.2 or higher as soon as possible to resolve this issue. This update will ensure enabled long-acting reminders are delivered as expected. For iOS Users: To confirm the latest app version on your iOS device, follow the steps below: 1. Open the App Store. 2. Tap your profile icon at the top of the screen. 3. Scroll to see pending updates. 4. Tap Update next to the MiniMed Go app icon. For Android Users: To confirm the latest app version on your Android device, follow the steps below: 1. Open the Google Play Store. 2. Tap Menu, then select Manage apps & device. 3. Apps with an update available are labeled Update available. You can also search for theMiniMed Go app. 4. Tap Update next to the MiniMed Go app icon. For Questions or assistance - call the Medtronic MiniMed 24-Hour Technical Support line at 1-800-646-4633, option 1.
- Initiated
- 2026-09-02
- Root cause
- Software design
- Distribution
- Worldwide - U.S. Nationwide distribution in the states of AL, CA, CT, FL, LA, MA, MD, NY, OH, PA, and TX. The countries of Austria, Belgium, Denmark, Finland, Germany, Iceland, Italy, Norway, Spain, Sweden, Switzerland, United Kingdom.
- Recall number
- Z-0265-2027
- Recall status
- Open, Classified
- CDRH record ID
- 222954
- FDA classified
- 2026-10-22
- Recall event number
- 99945