Open original FDA record ↗Recall record
- FDA identifier
- Z-0266-2008
- Company in FDA record
- Telefelx Medical
- FDA classification date
- Nov 29, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The specified insulation is not present on the shaft of the devices, leaving a risk of arcing and/or damage to surrounding tissue.
Source facts
- Action
- Consignees were notified by an Urgent Medical Device Recall letter on/about 11/1/2007. They were instructed to immediately cease use and distribution of the affected product and to return them to Teleflex Medical, Durham, NC. In addition, each were instructed to forward the letter if they had further distribution into institutional settings or to homecare patients and retrieve relevant product form those locations. An Acknowledgement & Stock Status Form was enclosed to be completed and faxed back to Teleflex Medical.
- Initiated
- 2007-10-31
- Root cause
- Process control
- Terminated
- 2009-08-31
- Distribution
- Worldwide: USA including states of AL, AR, AZ, CA, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY, NV, OH, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV, and countries of Canada, France New Zealand and Singapore.
- Recall number
- Z-0266-2008
- Recall status
- Terminated
- CDRH record ID
- 65715
- FDA classified
- 2007-11-29
- Recall event number
- 45515