Z-0266-2022: Cobas u 601 urinalysis test system, Catalog number 06334601001 The cobas U 601 urinalysis test system is

Recall record Z-0266-2022 for Roche Diagnostics Operations, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0266-2022
Company in FDA record
Roche Diagnostics Operations, Inc.
FDA record creation date
Nov 18, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A potential risk for false negative nitrite results exists when endogenous creatinine levels are 15,000 mg/L and above. No interference of Nitrite results was observed at creatinine levels up to 9,000 mg/L.

Source facts

Action
On October 18, 2021, the firm, Roche, sent "URGENT MEDICAL DEVICE CORRECTION" letters to affected consignees. Customers were notified of the potential for false negative nitrite test results with elevated endogenous creatinine concentrations. Customers were provided with an amended cobas u pack method sheet that provides the updated interference claim for endogenous creatinine. Nitrite testing is commonly included in the process of diagnosing urinary tract infections. In case of false negative nitrite results, particularly in asymptomatic patients or patient with nonspecific symptoms, further diagnostic measures in urine might be skipped, resulting in an increased potential risk for disease progression and complications of UTI. Actions required: - Please consult with the physician or pathologist at your facility to determine specific clinical implications for your patients. - Review the updated cobas u pack method sheet included with the letter and update your laboratory's procedures, if needed. - Complete and return the customer response form via email to roche6051@sedgwick.com. - File the recall notice for future reference. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.
Initiated
2021-10-18
Root cause
Software design
Terminated
2023-05-31
Distribution
US Nationwide distribution.
Recall number
Z-0266-2022
Recall status
Terminated
CDRH record ID
189933
FDA record created
2021-11-18
Recall event number
88879

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗